New to system suitability testing? This guide explains the concept in plain language before you dive into the calculator.
System Suitability Testing verifies that an analytical system such as HPLC or UV-Vis is performing correctly before sample analysis begins. A standard solution is injected multiple times, typically 5–6 replicates, and the precision of those replicate responses is evaluated using %RSD and %CV.
Under USP <621>, the standard acceptance limit for %RSD in assay-type methods is typically 2.0% or less. Tighter limits of 1.0% or looser limits up to 5.0% may apply depending on the specific method SOP and test type.
Yes. %RSD (relative standard deviation) and %CV (coefficient of variation) are numerically identical. Both are calculated as the standard deviation divided by the mean, multiplied by 100, and both measure the relative spread of a dataset.
System suitability is checked at the start of every analytical run before sample injections, and it is required under USP <621>, ICH Q2(R1), and FDA guidance. It is also re-verified periodically during long sequences using bracketing standards to confirm the system remains stable.